Hello , Sign In
The incubator Im is a Class I medical device.
Product Type | Incubator |
---|---|
Alarm | Visual and Acoustic |
Altitude of Installation | Max. 2,000 M Above Sea Level |
Ambient Temperature | +5 °C to +40 °C |
Autodiagnostic System | For Fault Analysis |
Calibration | Three Freely Selectable Temperature Values |
Controlcockpit | Twindisplay. Adaptive Multifunctional Digital PID-Microprocessor Controller With 2 High-Definition TFT-Colour Displays. |
Timer Range | 1 min to 99 days |
Chamber Volume | 108 L |
Convection | Natural Convection |
Dimensions | 745 x 864 x 584 mm |
Interior Dimensions | 560 x 480 x 400 mm |
Exterior Dimensions | 745 x 864 x 584 mm |
Volume | 108 L |
Voltage Rating | 115 V;230V |
Door | Fully Insulated Stainless Steel Door With 2-Point Locking (compression Door Lock)~Inner Glass Door |
Frequency Rating | 50/60 Hz |
Fresh Air | Admixture of Pre-Heated Fresh Air By Electronically Adjustable Air Flap |
Number of Shelves | 2 (Standard);5 (Maximum) |
Function Heatbalance | Adapting The Distribution of The Heating Performance of The Upper and Lower Heating Circuit From -50 % to +50 % |
Housing | Rear Zinc-Plated Steel |
Standards Met | CE Certified;cUL US Listed;DVE |
Power Rating | 900 W;1400W |
Humidity RH | Max. 80 %, Non-Condensing |
Interior | Easy-To-Clean Interior,Made of Stainless Steel, Reinforced By Deep Drawn Ribbing With Integrated and Protected Large-Area Heating on Four Sides |
Material | Zinc-Plated Steel |
Relative Humidity | Maximum 80 (Non-Condensing) % |
Internals | 2 Stainless Steel Grid(S), Electropolished |
Language Setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Maximum Loading of Chamber | 175 Kg |
Maximum Loading Per Internal | 20 Kg |
Maximum Loading of Internal | 20 kg |
Maximum Number of Internals | 5 |
Weight | 76 kg |
Net Weight | Approx. 76 Kg |
Software Requirement | AtmoCONTROL |
Operating Temperature Range | 5 to 80 °C |
Overvoltage Category | II |
Pollution Degree | 2 |
Programming | Atmocontrol Software on a USB Stick for Programming, Managing and Transferring Programmes Via Ethernet Interface or USB Port |
Set Up | The Distance Between The Wall and The Rear of The Appliance Must Be at Least 15 Cm. The Clearance From The Ceiling Must Not Be Less Than 20 Cm and The Side Clearance From Walls or Nearby Appliances Must Not Be Less Than 5 Cm. |
Setting Accuracy Temperature | 0.1 °C |
Setting Temperature Range | +20 to +80 °C |
Sterilisation | Fixed Sterilisation Programme (4 Hours/160°C) for Sterilisation of Working Chamber, Not for Sterilising The Load |
Temperature Sensor | 2 Pt100 Sensors Din Class a In 4-Wire-Circuit for Mutual Monitoring, Taking Over Functions In Case of An Error |
Timer | Digital Backwards Counter With Target Time Setting, Adjustable From 1 Minute to 99 Days |
Vent | Vent Connection With Restrictor Flap |
Voltage Electrical Load | 230 V, 50/60 Hz ,1400 W~115 V, 50/60 Hz ,900 W |
Working Temperature Range | At Least 5 Above Ambient Temperature to +80 °C |
Works Calibration Certificate | Incl. Works Calibration Certificate for +37°C |
Adjustable Parameters | Temperature (Celsius or Fahrenheit), Air Flap Position, Programme Time, Time Zones, Summertime/Wintertime |
Creating an account has many benefits: