Product Type Incubator
Alarm Visual and Acoustic
Altitude of Installation Max. 2,000 M Above Sea Level
Ambient Temperature +5 °C to +40 °C
Autodiagnostic System For Fault Analysis
Calibration Three Freely Selectable Temperature Values
Timer Range 1 min to 99 days
Controlcockpit Twindisplay. Adaptive Multifunctional Digital PID-Microprocessor Controller With 2 High-Definition TFT-Colour Displays.
Chamber Volume 32 L
Dimensions 585 x 704 x 434 mm
Interior Dimensions 400 x 320 x 250  mm
Exterior Dimensions 585 x 704 x 434 mm
Door Fully Insulated Stainless Steel Door With 2-Point Locking (compression Door Lock)~Inner Glass Door
Voltage Rating 115 V;230V
Frequency Rating 50/60 Hz
Fan Forced Air Circulation By Quite Air Turbine, Adjustable In 10 % Steps for Each Segment Individually
Fresh Air Admixture of Pre-Heated Fresh Air By Electronically Adjustable Air Flap
Number of Shelves 1;3 (Maximum)
Weight 48 kg
Standards Met CE Certified;cUL US Listed;DVE
Housing Rear Zinc-Plated Steel
Humidity RH Max. 80 %, Non-Condensing
Power Rating 800 W;1600W
Interior Easy-To-Clean Interior,Made of Stainless Steel, Reinforced By Deep Drawn Ribbing With Integrated and Protected Large-Area Heating on Four Sides
Material Zinc-Plated Steel
Relative Humidity Maximum 80 (Non-Condensing) %
Internals 1 Stainless Steel Grid(S), Electropolished
Language Setting German, English, Spanish, French, Polish, Czech, Hungarian
Maximum Loading of Chamber 60 Kg
Maximum Loading of Internal 20 kg
Maximum Loading Per Internal 20 Kg
Maximum Number of Internals 3
Net Weight Approx. 48 Kg
Overvoltage Category II
Operating Temperature Range 10 to 80 °C
Pollution Degree 2
Set Up The Distance Between The Wall and The Rear of The Appliance Must Be at Least 15 Cm. The Clearance From The Ceiling Must Not Be Less Than 20 Cm and The Side Clearance From Walls or Nearby Appliances Must Not Be Less Than 5 Cm.
Setting Accuracy Temperature 0.1 °C
Setting Temperature Range +20 to +80 °C
Sterilisation Fixed Sterilisation Programme (4 Hours/160°C) for Sterilisation of Working Chamber, Not for Sterilising The Load
Temperature Sensor 2 Pt100 Sensors Din Class a In 4-Wire-Circuit for Mutual Monitoring, Taking Over Functions In Case of An Error
Timer Digital Backwards Counter With Target Time Setting, Adjustable From 1 Minute to 99 Days
Vent Vent Connection With Restrictor Flap
Voltage Electrical Load 230 V, 50/60 Hz ,1600 W~115 V, 50/60 Hz ,800 W
Volume 32 L
Working Temperature Range Min. 10°C Above Ambient Up to +80°C
Works Calibration Certificate Incl. Works Calibration Certificate for +37°C
Adjustable Parameters Temperature (Celsius or Fahrenheit), Fan Speed, Air Flap Position, Programme Time, Time Zones, Summertime/Wintertime