Maximum energy efficiency in continuous operation - tailor-made for stability studies according to ICH guidelines, stability tests for cosmetics and foods as well as environmental testing and material testing
Alarm | Visual and Acoustic |
---|---|
Product Type | Constant Climate Chamber |
Altitude of Installation | Max. 2,000 M Above Sea Level |
Chamber Volume | 1360 L |
Ambient Temperature | 16 °C to 28 °C |
Humidity Range | 70 % (Non-Condensing) |
Autodiagnostic System | Integral Fault Diagnostics for Temperature and Humidity Control |
Calibration | Three Freely Selectable Values Each, Temperature and Humidity |
Sensor Type | PT100 Sensor |
Interior Dimensions | 1250 x 1450 x 750 mm |
Dimensions | 1435 x 1913 x 905 mm |
Exterior Dimensions | 1435 x 1913 x 905 mm |
Weight | 347 kg |
Voltage Rating | 230 V;115V |
Frequency Rating | 50/60 Hz |
Door | Stainless Steel Doors With Glass Sectors, Fully Heated Inner Glass Panes Integrated In The Full-Sight Glass Door With 2-Point Locking (compression Door Lock) |
Standards Met | CE Certified |
Power Rating | 1400 W |
Maximum Loading of Internal | 30 kg |
Housing | Rear Zinc-Plated Steel |
Material | Zinc-Plated Steel |
Humidification | Humidification By Hot Steam Generator |
Humidity RH | Max. 70 %, Non-Condensing |
Installation | On Lockable Castors, Adjustable In Height |
Internals | 4 Stainless Steel Grid(S), Electropolished |
Maximum Loading Per Internal | 30 Kg |
Maximum Number of Internals | 28 |
Net Weight | Approx. 347 Kg |
Pollution Degree | 2 |
Programming | Atmocontrol Software on a USB Stick for Programming, Managing and Transferring Programmes Via Ethernet Interface or USB Port |
Set Up | The Distance Between The Wall and The Rear of The Appliance Must Be at Least 15 Cm. The Clearance From The Ceiling Must Not Be Less Than 20 Cm and The Side Clearance From Walls or Nearby Appliances Must Not Be Less Than 5 Cm. |
Setting Accuracy Temperature | 0.1 °C |
Temperature Sensor | 2 Pt100 Sensors Din Class a In 4-Wire-Circuit for Mutual Monitoring, Taking Over Functions In Case of An Error |
Timer | Digital Backwards Counter With Target Time Setting, Adjustable From 1 Minute to 99 Days |
Volume | 1360 L |
Controlcockpit | Twindisplay. Adaptive Multifunctional Digital PID-Microprocessor Controller With 2 High-Definition TFT-Colour Displays. |
Dehumidification | Dehumidification By Cold Trap Using The Peltier Technology |
Function Heatbalance | Adapting The Distribution of The Heating Performance of The Upper and Lower Heating Circuit From -50 % to +50 % |
Humidity | Humidity Supply With Distilled Water From External Tank By Self-Priming Pump |
Interface | Ethernet LAN, USB |
Language Setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Maximum Loading of Chamber | 250 Kg |
Overvoltage Category | II |
Peltier | Energy-Saving Peltier Heating-/Cooling System Integrated In The Rear (heat Pump Principle) |
Setting Accuracy Humidity | 0.5 % Rh |
Setting Range Humidity | 10 - 80 % Rh |
Setting Temperature Range | Without Light, With Humidity: +15 to +60 °C~without Light, Without Humidity: 0 to +70 °C |
Standard Accessories | Water Tank Including Connection Hose |
Standard Works Calibration Certificate | +25 °C / 60 % Rh, +40 °C / 75 % Rh |
Voltage Electrical Load | 230 V, 50/60 Hz ,1400 W~115 V, 50/60 Hz ,1400 W |
Working Temperature Range | Without Light, With Humidity: +15 (at Least 10 Below Ambient Temperature) to +60 °C~without Light, Without Humidity: 0 (at Least 20 Below Ambient Temperature) to +70 °C |
Adjustable Parameters | Temperature (Celsius or Fahrenheit), Relative Humidity, Programme Time, Time Zones, Summertime/Wintertime |