Hello , Sign In
Safety at all times: High-end functions for the protection of cell cultures, bacteria cultures or tissue cultures.
Adjustable Parameters | Temperature (Celsius or Fahrenheit), CO2, Programme Time, Time Zones, Summertime/Wintertime |
---|---|
Product Type | CO2 Incubator |
Adjustment Range CO2 | 0 to 20 % CO2 |
Adjustment Range O2 | 1 to 20 % O2 |
Alarm | Visual and Acoustic |
Altitude of Installation | Max. 2,000 M Above Sea Level |
Ambient Temperature | 10 °C to 35 °C |
Autodiagnostic System | Integral Fault Diagnostics for Temperature and CO2 |
Chamber Volume | 107 L |
CO2 Control | Digital Electronic CO2 Control With Dual Beam Ndir System, With Auto-Diagnostic System and Acoustic Fault Indication, Barometric Pressure Compensation |
Controlcockpit | Twindisplay. Adaptive Multifunctional Digital PID-Microprocessor Controller With 2 High-Definition TFT-Colour Displays. |
Interior Dimensions | 560 x 480 x 400 mm |
Exterior Dimensions | 719 x 850 x 591 mm |
Volume | 107 L |
Voltage Rating | 115 V;230V |
Dimensions | 719 x 850 x 591 mm |
Frequency Rating | 50/60 Hz |
Number of Shelves | 2 (Standard);6 (Maximum) |
Door | Fully Insulated Stainless Steel Door With 2-Point Locking (compression Door Lock)~Inner Glass Door With Opening (8 mm Ø) to Take Gas Sample |
Relative Humidity | Maximum 70 (Non-Condensing) % |
Standards Met | CE Certified;cUL US Listed;DVE GS |
Housing | Rear Zinc-Plated Steel |
Humidity Control | Humidity Limitation Thanks to a Peltier Element; When Water Dish Is Full and Inserted, The Peltier Element Limits The Value of Relative Humidity In The Interior to 93 % Rh +/- 2.5 % |
Power Rating | 1300 W |
Material | Zinc-Plated Steel |
Humidity RH | Max. 70 %, Non-Condensing |
Interface | Ethernet LAN, USB |
Interior | Material 1.4301 (ASTM 304), Corrosion Resistant |
Language Setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Maximum Loading of Internal | 15 kg |
Maximum Loading of Chamber | 90 Kg |
Maximum Loading Per Internal | 15 Kg |
Software Requirement | AtmoCONTROL |
Weight | 75 kg |
Operating Temperature Range | 5 to 50 °C |
Maximum Number of Internals | 6 |
Net Weight | Approx. 75 Kg |
Overvoltage Category | II |
Pollution Degree | 2 |
Programming | Atmocontrol Software on a USB Stick for Programming, Managing and Transferring Programmes Via Ethernet Interface or USB Port |
Set Up | The Distance Between The Wall and The Rear of The Appliance Must Be at Least 15 Cm. The Clearance From The Ceiling Must Not Be Less Than 20 Cm and The Side Clearance From Walls or Nearby Appliances Must Not Be Less Than 5 Cm. |
Setting Accuracy CO2 | 0,1% |
Setting Accuracy O2 | 0.1 % O2 |
Setting Accuracy Humidity | 0.5 % Rh |
Setting Accuracy Temperature | 0.1 °C |
Setting Range Active Humidity Control (with Option K7) | 40 to 97 % Rh and Rh-Off |
Setting Temperature Range | +18 to +50 °C |
Standard Accessories | Membrane Filter (in Order to Remove Impurities and Pollutants, All Incoming Gases Pass Through a Membrane Filter Before They Reach The Chamber) |
Standard Equipment | CO2 Connection Set: Hose With Coupling and Clamp |
Temperature Fluctuations With Time | At 37 °C +/- 0.1 K |
Temperature Sensor | 2 Pt100 Sensors Din Class a In 4-Wire-Circuit for Mutual Monitoring, Taking Over Functions In Case of An Error |
Temperature Variation In Chamber | At + 37 °C +/- 0.3 K |
Variation In Time CO2 | +/- 0.2 % CO2 |
Voltage Electrical Load | 230 V, 50/60 Hz ,1300 W~115 V, 50/60 Hz ,1300 W |
Working Temperature Range | 5 °C Above Ambient Temperature Up to +50 °C Standard Sterilisation Programme: 60 Minutes at 180°C (without Removing The Sensors) |
Works Calibation Certificate | Works Calibration Certificate at +37 °C, 5 % CO2 and 90 % Rh |
Internals | 2 Perforated Stainless Steel Shelf/Shelves~1 Stainless Steel Water Dish |
Creating an account has many benefits: