Unparalleled homogeneity of temperature and humidity. For stability testing according to ICH - WHO - EMA - ASEAN - GMP - GLP – GCCP.
Alarm | Visual and Acoustic |
---|---|
Product Type | Climate Chamber |
Altitude of Installation | Max. 2,000 M Above Sea Level |
Chamber Volume | 108 L |
Ambient Temperature | 15 °C to 28 °C (up to 34 °C With Limited Temperature Range) |
Humidity Range | 70 % (Non-Condensing) |
Autodiagnostic System | Integral Fault Diagnostics for Temperature and Humidity Control |
Calibration | Three Freely Selectable Values Each, Temperature/Humidity/CO2 |
Cooling | Energy-Saving Use of Cfc-Free Cooling/Heating System (refrigerant R134a) |
Sensor Type | PT100 Sensor |
Interior Dimensions | 560 x 480 x 400 mm |
Dimensions | 745 x 1233 x 585 mm |
Exterior Dimensions | 745 x 1233 x 585 mm |
Weight | 109 kg |
Voltage Rating | 230 V;115V |
Frequency Rating | 50/60 Hz |
Door | Fully Insulated Stainless Steel Door With 2-Point Locking (compression Door Lock)~Inner Glass Door |
Standards Met | CE Certified;EAC |
Entry Port | Entry Port (silicone), 40 mm Clear Diameter, Moisture Tight, Can Be Closed By a Silicone Stopper, Standard Position at The Back |
Power Rating | 1350 W |
Fan | Forced Air Circulation In Working Chamber Adjustable From 10 to 100 % In 10 % Steps |
Maximum Loading of Internal | 20 kg |
Housing | Rear Zinc-Plated Steel |
Material | Zinc-Plated Steel |
Humidification | Humidification By Hot Steam Generator |
Humidity RH | Max. 70 %, Non-Condensing |
Installation | On Lockable Castors |
Internals | 2 Stainless Steel Grid(S), Electropolished |
Maximum Loading Per Internal | 20 Kg |
Maximum Number of Internals | 5 |
Net Weight | Approx. 109 Kg |
Pollution Degree | 2 |
Programming | Atmocontrol Software on a USB Stick for Programming, Managing and Transferring Programmes Via Ethernet Interface or USB Port |
Set Up | The Distance Between The Wall and The Rear of The Appliance Must Be at Least 15 Cm. The Clearance From The Ceiling Must Not Be Less Than 20 Cm and The Side Clearance From Walls or Nearby Appliances Must Not Be Less Than 5 Cm. |
Setting Accuracy Temperature | 0.1 °C |
Temperature Sensor | 2 Pt100 Sensors Din Class a In 4-Wire-Circuit for Mutual Monitoring, Taking Over Functions In Case of An Error |
Timer | Digital Backwards Counter With Target Time Setting, Adjustable From 1 Minute to 99 Days |
Volume | 108 L |
Works Calibration Certificate | Standard Values +10 °C and +37 °C, 60 % Rh at +30 °C |
Controlcockpit | Twindisplay. Adaptive Multifunctional Digital PID-Microprocessor Controller With 2 High-Definition TFT-Colour Displays. |
Humidity | Humidity Supply With Distilled Water From External Tank By Self-Priming Pump~Active Humidifying and De-Humidifying Adjustable From 10-80 % Rh With Digital Display of Relative Humidity - Resolution of Display 0.1 %, Setting Accuracy 0.5 % |
Language Setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Maximum Loading of Chamber | 150 Kg |
Overvoltage Category | II |
Setting Temperature Range | -10 to +60 °C |
Standard Accessories | Water Tank Including Connection Hose |
Voltage Electrical Load | 230 V, 50/60 Hz ,1350 W~115 V, 50/60 Hz ,1350 W |
Working Temperature Range | From -10°C Up to +60°C Without Humidity (not Suitable for Long-Term Storing at Sub-Zero Temperatures. During Permanent Operation, The Glass Door May Ice Over)~From +10°C Up to +60°C With Humidity And/Or Light |
Adjustable Parameters | Temperature (Celsius or Fahrenheit), Fan Speed, Relative Humidity, Programme Time, Time Zones, Summertime/Wintertime |
Air Jacket | Air Jacket Heating System for Gentle All-Around Heating |
Defrosting | Highly Efficient Automatic Defrosting System |
Heating Concept | No Drying-Up of The Load Caused By The Cooling Device Due to Separation of Thermal Jacket From Interior |