Hello , Sign In
The outstanding success of the Polilyte Plus in chemical and wastewater applications gave the inspiration for transferring the good features to a whole family of sensors. The expanded portfolio widens the range of applications that can be covered
ATEX Approval | II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db |
---|---|
Autoclavable | Yes |
CIP | No |
Diameter | 12 mm |
Diaphragm | Single Pore |
Electrical Connector | S8 |
Electrolyte | Polisolve Plus |
IECEx Approval | II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db |
Liquid Earth | No |
Measurement Principle | Combination electrode; pH potential measured against reference |
Measuring Range | pH 0 to 14 |
Membrane Shape | Spherical |
Number of Diaphragms | 2 |
Operating Temperature Range | 0 to 130 °C |
Parameter | pH |
Pressure Range | 0 to 50 bar (60°C), 0 to 20 bar (100 °C), 0 to 16 bar (130°C) |
Process Connection | PG13,5 |
Recommended Storage | Hamilton Storage Solution or 3 M KCl |
Reference System | Hamilton Everef-L |
Sample min Conductivity | 2 µS/cm |
Sensitivity | 57 to 59 mV / pH at 25 °C |
Sensor Family | Polilyte Plus |
Sensor Output | Traditional: mV |
Serial Number | Yes |
Steam Sterilizable | Yes |
Temperature Sensor | No |
Wetted Parts | Glass \nFKM (Fluorocarbon Elastomer) |
Zero Point | 0 ± 20 mV |
Length | 120 mm |
pH Glass | Hamilton Type H Glass |
pH Glass Resistance | Moderate |
O-ring Material | FDA 21 CFR 177.2600; EG 1935/2004; USP <87>; USP <88> Class VI (121 °C) |
Creating an account has many benefits: