The IonoTrode sensor is designed for applications in ion weak media. The F glass membrane has a very low resistance, therefore the sensor can be used in samples with low conductivity, where it offers highest accuracy over a long period of time
CIP | No |
---|---|
Diameter | 12 mm |
Diaphragm | Sleeve |
Electrical Connector | S7 |
Electrolyte | 3 M KCl |
Liquid Earth | No |
Measurement Principle | Combination electrode; pH potential measured against reference |
Measuring Range | pH 0 to 14 |
Membrane Shape | Spherical |
Operating Temperature Range | -10 to 40 °C |
Pressure Range | 0 to 0.5 bar (higher if side-arm is pressurized) |
Recommended Storage | Hamilton Storage Solution or 3 M KCl |
Reference System | Hamilton Everef |
Sample min Conductivity | 0.2 µS/cm |
Sensitivity | 57 to 59 mV / pH at 25 °C |
Sensor Family | IonoTrode |
Sensor Output | Traditional: mV |
Serial Number | Yes |
Steam Sterilizable | No |
Temperature Sensor | No |
Wetted Parts | Glass \nEPDM (Ethylene propylene elastomer) |
Zero Point | 0 ± 20 mV |
Length | 120 mm |
pH Glass | Hamilton Type F Glass |
pH Glass Resistance | Low |
O-ring Material | FDA 21 CFR 177.2600; EG 1935/2004; USP <87>; USP <88> Class VI (121 °C) |