Hello , Sign In
The InchTrode sensors are designed to measure pH in demanding applications in the paper making as well as in the chemical industries. The Single Pore liquid junction guarantees the best and fast measuring results because of direct contact between the sample and the Polisolve electrolyte
ATEX Approval | II 1/2 G Ex ia IIB T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db |
---|---|
Autoclavable | No |
CIP | No |
Diameter | 22 mm |
Diaphragm | Single Pore ring |
Electrical Connector | 10 m Fix Cable / Open End |
Electrolyte | Polisolve |
IECEx Approval | II 1/2 G Ex ia IIB T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db |
Liquid Earth | Yes |
Measurement Principle | Combination electrode; pH potential measured against reference |
Measuring Range | pH 0 to 14 |
Membrane Shape | Flat |
Number of Diaphragms | 1 |
Operating Temperature Range | -10 to 130 °C |
Parameter | pH |
Pressure Range | 0 to 10 bar at 25 °C, 0 to 6 bar at 130 °C |
Process Connection | NPT ¾" Thread |
Recommended Storage | Hamilton Storage Solution or 3 M KCl |
Reference System | Hamilton Everef-L |
Sample min Conductivity | 2 µS/cm |
Sensitivity | 57 to 59 mV / pH at 25 °C |
Sensor Family | InchTrode |
Sensor Output | Traditional: mV |
Serial Number | Yes |
Steam Sterilizable | Yes |
Temperature Sensor | Pt100 |
Wetted Parts | Glass \nPEEK (Polyetheretherketone) |
Zero Point | 0 ± 20 mV |
Length | 143 mm |
pH Glass | Hamilton Type HF Glass: For < 0.01M / 200 mg/l HF at 20 °C or 0.05M / 1000mg/l HF at 50 °C |
pH Glass Resistance | Moderate |
Creating an account has many benefits: