Hello , Sign In
This unit is used in liquid chromatography to verify analyzed samples by means of eight wavelengths or scanned spectra. Noise level at 254 nm is ±5·10-6 AU with a test cell. High standard deuterium lamp in a special socket enables easy exchange from front panel. Output signal is available in both digital and analogue form. Detector could be controlled by Clarity or ECOMAC software. Cell could be easily removed from the front of the unit. Three selectable flow cells AD02, AD05 and AD10 with optical path length 2.4, 5 and 10 mm. AD05 cell is included as component of the detector. Other cells on request
Wavelength Range | 200-800 nm |
---|---|
Accuracy of Adjustment | ± 1 nm |
Reproducibility | ± 0.5 nm |
Light Source | Deuterium Lamp, Halogen Lamp |
Noise | ± 5 x 10-6 AU |
Drift | 1·10-4 AU/hr |
Time Constant | 10 - 10 000 ms |
Sampling Speed | up to 100 Hz |
Digital output | 1 V/AU |
Analog Output | 4x Configurable |
Wetted Materials | Fused Silica, Stainless Steel, PEEK |
Communication | RS232,Ethernet (LAN),USB |
Display | VFD 140x32 Pixels |
Keypad | 10 pushbuttons |
Power Rating | 110 VA |
Voltage Rating | 100-240 V |
Frequency Rating | 50/60 HZ |
Weight | 9 kg |
Optical Cables | No |
Number of CCD Elements | 512 Mathematically Enhanced to 4 096 by Sophisticated Algorithm |
Dimensions (W X H X D) | 280 x 135 x 498 mm (11.0”x 5.3” x 19.6”) |