Hello , Sign In
The B310 is an addition to the new B300 system, which produces all types of water for laboratory applications: Grade 1, Grade 2 as well as reverse osmosis water. Adrona B310 is ultrapure water purification system which requires external water tank that use tap water as feed water (tap water system). There are two variations available: B310 HPLC and B310 Bio. The B310 has completely new electronics and software, providing a number of additional features: large colour touch screen, data logging capability, warning and alarm messages, a wide range of flow control options, the ability to connect multiple remote dispensers, Ethernet and USB interfaces, dispense report preparation, improved accuracy of conductivity and TOC measurements.
Product Type | Water Purification System |
---|---|
Model | HPLC |
Grade 1 Water Resistivity at 25 °C | 18.2 MΩ x cm |
Grade 1 Water Conductivity At 25°C | 0.055 μS/cm |
Grade 2 Water Resistivity at 25 °C | >10 MΩ x cm |
Grade 2 Water Conductivity at 25 °C | <0.1 μS/cm |
TOC Level | < 5 ppb |
Bacteria | < 0.01 CFU/mL |
Endotoxins | < 0.15 EU/mL |
Particles >0.22 μm | < 1/mL |
Dispense Rate, Ultrapure Water | 1.5 to 2 L/min |
Dimensions (WxDxH) | 35x39x54 cm |
System Weight | 28 kg |
Operating Weight | 31 kg |
Volumetric Dispense | 0.01 to 100 L |
Feed Water | Potable Tap Water |
Conductivity | <2000 μS/cm |
Pressure | 0.4 to 6.0 bar |
Temperature | 5 to 35 °C |
Dispenser | Dispense Port |
Mains Voltage | 200 to 240 VAC |
Power | 130 W |
Fuse | 230 V, 3.15 A |
Tank | Required, Not Included |
pH | 4 to 10 |
Fouling Index | <5 |
Iron | <0.1 ppm as CaCO3 |
Aluminium | <0.05 ppm as CaCO3 |
Manganese | <0.05 ppm as CaCO3 |
Total Chlorine | <1 ppm |
Langelier Saturation Index | <+0.2 |
Including | water storage tank "Pro", external prefilter, AquaStop and CO2 tank filter |
Relative Humidity | 20–80% |
Nominal Flow | 10 L/h |