Hello , Sign In
Adrona B300 is ultrapure water purification system which requires external water tank that use tap water as feed water (tap water system). There are 3 variations available: B300 Trace, B300 HPLC, B300 Bio. The B300 is a new integrated Grade 2/ Grade 1 laboratory water system, replacing the customer-approved previous B30 model. The B300 has completely new electronics and software, providing a number of additional features: large colour touch screen, data logging capability, warning and alarm messages, a wide range of flow control options, the ability to connect multiple remote dispensers, Ethernet and USB interfaces, dispense report preparation, improved accuracy of conductivity and TOC measurements.
Product Type | Water Purification System |
---|---|
Model | HPLC |
Grade 1 Water Resistivity at 25 °C | 18.2 MΩ x cm |
Grade 1 Water Conductivity At 25°C | 0.055 μS/cm |
Grade 2 Water Resistivity at 25 °C | >10 MΩ x cm |
Grade 2 Water Conductivity at 25 °C | <0.1 μS/cm |
TOC Level | < 5 ppb |
Bacteria | < 0.01 CFU/mL |
Endotoxins | < 0.15 EU/mL |
Particles >0.22 μm | < 1/mL |
Dispense Rate, Ultrapure Water | 1.5 to 2 L/min |
Dimensions (WxDxH) | 35x39x54 cm |
System Weight | 28 kg |
Operating Weight | 31 kg |
Volumetric Dispense | 0.01 to 100 L |
Nominal flow, pure water (to storage tank) | 10 L/h |
Feed Water | Potable Tap Water |
Conductivity | <2000 μS/cm |
Pressure | 0.4 to 6.0 bar |
Temperature | 5 to 35 °C |
Dispenser | Dispense Port |
Mains Voltage | 200 to 240 VAC |
Power | 130 W |
Fuse | 230 V, 3.15 A |
Tank | Required, Not Included |
pH | 4 to 10 |
Fouling Index | <5 |
Iron | <0.1 ppm as CaCO3 |
Aluminium | <0.05 ppm as CaCO3 |
Manganese | <0.05 ppm as CaCO3 |
Total Chlorine | <1 ppm |
Langelier Saturation Index | <+0.2 |
Including | water storage tank "Pro", external prefilter, AquaStop and CO2 tank filter |
Relative Humidity | 20–80% |